Accel-RF offers a complete lab planning guide, specific to each test system sold.
For general guidelines, please consult the following table:
Additional requirements:
Electrical: (should have backup UPS or other preventive power loss protection)
120/208VAC 50A (dependent on DUT/SSPA ratings), 3-Phase, 5-Wire, Wye configurationor
230/400VAC 30A (dependent on DUT/SSPA ratings), 3-Phase, 5-Wire, Wye configuration
Step up Transformer (208/400VAC) with 20KVA load capacity.
TBD Plug and Receptacle (dependent on current rating)
Size:
22.5”W x 58”H x 36”D (single-bay)
46”W x 58”H x 36”D (dual-bay)
68”W x 58”H x 36”D (triple-bay)
Approximately 750lbs. (single-bay) / 1200lbs. (dual-bay) / 1500lbs. (triple-bay)
Chiller Size (if applicable):
22.5”W x 22”H x 42”D (est.)
Chilled Water (if applicable):
Chiller with 7,000W capacity capable of maintaining 12°C
Or
Facility water (filtered) with temperature of 12°C and 50psi pressure (nominal values)
Environmental:
Room Temperature should be +23oC +/- 5oC Humidity: standard room environmentArea: at least 1 meter clearance at back and front of system